Fluke Biomedical 8000 Victoreen User Manual
Page 57

Operation
Modes of Operation
2
2-45
To Set Air Density (Temperature & Pressure)
SETUP
CLOCK
− >
OFF
CAL
− >
OFF
HVL
UNITS
− >
R
EXP
AIR
−−− >
20.5 C
734 mmHg
CT EXP
DATE
−− >
Oct. 10,
1996
AMSE
TIME
−−>
10 30 45
MODE SELECT
TEMP.
PRESSURE
From the SETUP screen, use the select softkey (1) to select AIR. When AIR is selected, the air density
selection blinks and a highlight (reverse video) extends across the other 2 display fields. The temperature
and pressure keys select and increment temperature and pressure. Holding down either of these keys
will continuously increment the selected value. Temperature and pressure may also be incremented or
decremented with the up or down keys. Once the desired temperature and pressure have been set, press
the ENTER key to accept them. Default values for temperature and pressure are 22° C and 760 mmHg.
(Refer to Appendix A--Temperature and Pressure--for additional information.)
To Select Exposure Units (R or Gy)
SETUP
CLOCK
− >
OFF
CAL
− >
OFF
HVL
UNITS
− >
R
EXP
AIR
−−− >
20.5 C
734 mmHg
CT EXP
DATE
−− >
Oct. 10,
1996
AMSE
TIME
−−>
10 30 45
MODE SELECT
R/Gy
From the SETUP screen, use the select softkey (1) to select UNITS. When UNITS is selected, the units
selection blinks and a highlight (reverse video) extends across the other display field. The R/Gy softkey
(2) toggles the exposure units between R (Roentgens) and Gy (Grays). Exposure units may also be
toggled with the up or down keys.
To Turn Automatic Printing On or Off
SETUP
CLOCK
− >
OFF
CAL
− >
ON
HVL
UNITS
− >
R
EXP
AIR
−−− >
20.5 C
734 mmHg
CT EXP
DATE
−− >
Oct. 10,
1996
AMSE
TIME
−−>
10 30 45
MODE SELECT
ON/OFF
From the SETUP screen, use the select softkey (1) to select PRINT. When PRINT is selected, the print
selection blinks and a highlight (reverse video) extends across the other display field. The ON/OFF
softkey (2) toggles automatic printing between on or off. Automatic printing print may also be turned on or
off with the up or down keys.