5 modes of operation – Fluke Biomedical 8000 Victoreen User Manual
Page 22

Victoreen 8000
Operators Manual
2-10
Wait a moment while the 8000 analyzes the exposure then displays the measured kVp, exposure, time
and mAs (mAs will be zero because the mAs input was not used).
80.0
kVp Avg
100
msec
79.2
kV Eff
392
mR
81.1
kV Peak
0.0
mAs
RADIO
75%
LOW
+ 10 ms
Please
MODE %kV
SENS
DELAY
Wait…
The 8000 will immediately display Please Wait… in the lower right corner of its screen while it prepares to
take another exposure.
After the 8000 beeps and displays Make Exposure in the lower right corner of the display, another
exposure may be made.
To stop making exposures and return to the main menu screen, press the MODE button.
2.5 Modes of Operation
The NERO mAx Nero has ten modes of operation as listed below (in selection order):
1. RADIO
UNIT
2. MAMMO
SETUP
CAL
3. FLUORO
HVL
EXP
4. AMSE
CT EXP
1Ø PULSE
5. CT
EXP
AMSE ZERO
6. EXP
FLUORO
75%
MAMMO 80%
7. HVL
LOW
RADIO
90%
8. CAL
HIGH
10 mS
MODE %kV
9. SETUP
SENS DELAY
10. UNIT ID
The NERO mAx's mode of operation is controlled from the main screen. When the menu cursor (blinking
reverse video) is moved to a mode selection, by pressing the mode key or using the up and down arrow
keys, all of the options available for that mode are displayed. Pressing a softkey (1 thru 4) moves the
menu cursor into the selected menu column and starts it blinking. Pressing the softkey again will move
the blinking menu cursor up through the column of available options. Pressing the up and down arrow
keys moves the blinking menu cursor up or down through the selected column of available options. After
options have been highlighted, the enter key is pressed to start the selected mode with the selected
options. Pressing the mode key at any time returns to the mode select screen.
NERO mAx Operating Modes
Radio Mode
Radio mode is used to make measurements on tungsten target, aluminum filtered radiographic x-ray
generators. Radio mode simultaneously measures kVp, exposure, exposure time and mAs from a single
radiographic exposure. Radiographic mode may also be used to verify the NERO mAx’s kV calibration.