Caution Summary
The following CAUTIONS are provided for your reference and may appear throughout the NERO mAx
manual:
NOTE
CAUTION
CAUTION
Use extreme caution when connecting to the mAs
terminal of the detector.
If line voltage surges beyond 15% of normal, a
power line conditioner must be used, otherwise
damage to the charging circuit will occur.
Note Summary
In the event of a transient induced lockup of the
Model 8000 NERO mAx, it is necessary to reset the
unit by cycling its power (turning it off then on).
After reset, the unit will power up in its normal
operating mode.
1.8 Receiving Inspection
Upon receipt of the package:
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier
and notify Fluke Biomedical at 440.248.9300.
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good order.
1.9 Storage
If the unit is to be stored prior to use, pack it in the original container, if possible, and store in an
environment free of corrosive materials, fluctuations in temperature and humidity, and vibration and
shock.
Prior to use, check the condition and functionality of the device. Also check that the calibration is still
valid. Periodic recalibrations are usually required by individual radiation safety and/or quality assurance
programs. Please consult your local radiation safety or quality assurance office if you have any
questions.