Fluke Biomedical 8000 Victoreen User Manual
Page 36

Victoreen 8000
Operators Manual
2-24
318
μR/pulse
0
mAs/pulse
0.28
R/min
FLUORO PULSED LOW
MODE TYPE SENS
SAVE
After the exposure starts, the NERO mAx displays exposure rate (R/min, mR/pulse) and mAs/pulse.
When the desired exposure rate and/or mAs/pulse are displayed, press the SAVE key to acquire kV data.
318
μR/pulse
0
mAs/pulse
0.28
R/min
FLUORO PULSED LOW
TURN OFF
MODE TYPE SENS
BEAM
After kV data is acquired, the NERO mAx prompts to turn off the beam.
318
μR/pulse
0
mAs/pulse
0.28
R/min
FLUORO PULSED LOW
ANALYZE
MODE TYPE SENS
DATA
After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.
318
μR/pulse
0
mAs/pulse
0.28
R/min
FLUORO PULSED LOW
Please
MODE TYPE SENS
Wait…
After data analysis is complete, kV, exposure rate and mAs/pulse are displayed. If an overrange is
detected an error message is displayed. If the calculated kV is above or below the selected filter kV
range, “High” or “Low” is displayed instead of kV (see Section 4.3--Error Messages--for more information).
After the exposure data is displayed, the NERO mAx displays a "Please wait..." message as it prepares
for the next exposure.