Defib pulse repetition test – Fluke Biomedical 6000D Impulse User Manual
Page 91

Defibrillator/Transcutaneous Pacemaker Analyzer
Using Battery Performance Test Elements
77
Defib Pulse Repetition Test
The Defib Pulse Repetition test is used to determine if a
battery-powered defibrillator can deliver a specified
number of discharges within a specified amount of time.
Tables 26 and 27 list the Defib Pulse Repetition test
measurements and custom parameters.
Table 26. Defib Pulse Repetition Test Measurements
Measurement
Unit of Measure
Description
Energy
Joules
Amount of energy discharged by the defibrillator
The test stores multiple energy measurements – one per discharge.
Table 27. Defib Pulse Repetition Test Custom Parameters
Parameter Description
Preset Energy
The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
Number of Pulses
The number of discharges to be completed during the test
Number of Minutes
The maximum time allowed for this test to be completed