Setting the display contrast, Setting auto back light off, Setting up sound – Fluke Biomedical 6000D Impulse User Manual
Page 40: Displaying instrument information, Controlling the analyzer r, Controlling the analyzer remotely

Impulse 6000D, 7000DP
Users Manual
26
Setting the Display Contrast
The Analyzer’s display contrast can be set in one of two
ways. First, when the Analyzer displays Select a
device…, pressing
G
or
H
adjusts the display contrast.
Another method of adjusting contrast is through the
display setup menu. From the main setup menu, press the
softkey labeled Display. Next press the softkey labeled
Contrast. A scroll box opens above the softkey label
when contrast is adjusted by pressing
G
(darker) or
H
(lighter). Press the softkey labeled Done to save the
setting. This setting is now the value used when the
Analyzer is turned on.
Setting Auto Back Light Off
From the main setup menu, press the softkey labeled
Display. Next, press the softkey labeled Auto Back Light
Off. A scroll box opens at which point
G
or
H
will scroll
through Disabled, 30 seconds, and 60 seconds. When the
display settings are set, press the softkey labeled Done to
save the settings.
Setting Up Sound
The Analyzer’s internal beeper can be enabled or
disabled. When enabled, the volume can be set to low,
medium, or high. From the setup main menu, press the
softkey labeled Sound.
Next, pressing the softkey labeled Beeper simply toggles
the beeper on or off. Pressing the softkey labeled Volume
opens a scroll box above the softkey label. Use
G
or
H
to scroll through low, medium, and high volume settings.
Once the sound functions are set, press the softkey
labeled Done to store the settings.
Displaying Instrument Information
From the main setup menu, press the softkey labeled
More to reveal additional setup selections. Next, press the
softkey labeled Instrument Info to display the Analyzer’s
manufacturing date, firmware version, and serial number.
See Figure 29. Pressing the softkey labeled More displays
the last calibration date.
fak24.eps
Figure 29. Analyzer Information Screen
Controlling the Analyzer Remotely
Ansur test automation systems allow a solutions-based
approach to complete testing of the medical device under
test (DUT). Ansur helps you create standard work using
the test template/sequence (which is based on your