Fluke Biomedical 6000D Impulse User Manual
Page 75

Defibrillator/Transcutaneous Pacemaker Analyzer
Using Defibrillator Test Elements
61
Table 5. Energy Measurement Test Measurements
Measurement
Unit of Measure
Description
Energy
Joules
Amount of energy discharged by the defibrillator
Peak Voltage
Volts
Peak voltage detected during the discharge
Peak Current
Amperes
Peak current detected during the discharge
Pulse Width 50%
milliseconds
The width of the pulse at 50% of its peak
Pulse Width 10%
milliseconds
The width of the pulse at 10% of its peak
Table 6. Energy Measurement Test Custom Parameters
Parameter Description
Preset Energy
The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
ECG Waveform
The ECG waveform that the Analyzer should simulate during the test