Charge time test – Fluke Biomedical 6000D Impulse User Manual
Page 76

Impulse 6000D, 7000DP
Users Manual
62
Charge Time Test
The Charge Time test measures how long a defibrillator
takes to charge up to a specified energy level. Typically,
this test uses as a parameter the maximum energy level
available to the defibrillator. Table 7 lists the
measurements taken for this test. Custom parameters
available for the Charge Time tests are listed in Table 8.
Note
This test element supports the Auto Advance
Test Guide Setting.
Table 7. Charge Time Test Measurements
Measurement
Unit of Measure
Description
Charge Time
seconds
How long it took for the defibrillator to charge to the preset energy specified.
Energy
Joules
Amount of energy discharged by the defibrillator
Peak Voltage
Volts
Peak voltage detected during the discharge
Peak Current
Amperes
Peak current detected during the discharge
Pulse Width 50%
milliseconds
The width of the pulse at 50% of its peak
Pulse Width 10%
milliseconds
The width of the pulse at 10% of its peak