Sensitivity Test

Automatic Interactive Threshold Detection

Compatible pacer rates .......................................30 to 120 PPM

ECG R wave:

Waveforms ..........................................................Square, Triangle, Sine
Width ...................................................................1 to 19 (by 1) ms

20 to 95 (by 5) ms
100 to 300 (by 25) ms

Accuracy .............................................................± 5% of setting
Amplitude ............................................................0.05 to 0.95 (by 0.05) mV

1.0 to 5.0 (by 0.5) mV

Accuracy .............................................................± 5% of setting

Refractory Period Tests

Paced Refractory Period

......................................20 to 500 ms

Sensed Refractory Period

....................................15 to 500 ms

Accuracy

................................................................±1 ms

Pacer pulse rate

....................................................20 to 200 PPM

ECG

Waveform............................................................Triangle wave
Pulse width..........................................................40 ms
Amplitude ............................................................1.0 mV