Fluke Biomedical 6000D Impulse User Manual
Page 17

Defibrillator/Transcutaneous Pacemaker Analyzer
Safety Information
3
•
Do not operate the Analyzer with the battery
eliminator connected, unless connected
directly to mains power. During battery
operation, completely remove the battery
eliminator/charger from both the Analyzer
and wall socket.
•
Do not connect the Analyzer to a patient or
equipment connected to a patient. The
Analyzer is intended for equipment
evaluation only and should never be used
in diagnostics, treatment or in any other
capacity where the Analyzer would come in
contact with a patient.
•
Observe all precautions noted by the
Device Under Test (DUT) equipment
manufacturer when analyzing the DUT.
•
Use extreme caution when working with
voltages above 30 volts.
•
Use the proper terminals, functions and
ranges for the test being performed.