Test load selections, Measurements – Fluke Biomedical 6000D Impulse User Manual
Page 52

Impulse 6000D, 7000DP
Users Manual
38
Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)
Test Load Selections
Defibrillator Input
Fixed Load ..........................................................50 Ω
Accuracy .............................................................
±
1 %, non-inductive (<2
μ
H)
Power Rating.......................................................10 defib pulses of 360 J every 5 minutes
Pacemaker Input
Variable Load ......................................................50 to 1500
Ω
in 50
Ω
steps
Accuracy .............................................................
±
1 %, non-inductive (<2
μ
H)
Power Rating.......................................................5 W (average), 40 W (peak) @ 1000 Ω
Measurements
Manufacturer Specific Algorithms (plus general purpose default algorithm selection)
GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)
Current
Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)