Page 89: Using battery performance test elements

Defibrillator/Transcutaneous Pacemaker Analyzer
Using Battery Performance Test Elements
75
ECG Noise Immunity Test
The ECG Noise Immunity test checks the ability of the
ECG monitor to reject AC line frequency noise. The
custom parameters for this test are listed in table 23.
Note
This is a visual / audible test; the Analyzer takes
no measurements during this test.
Table 23. ECG Noise Immunity Test Custom Parameters
Parameter Description
Noise Amplitude
Indicates the amplitude for the waveform
Line Frequency
Selects 50 Hz or 60 Hz line noise simulation
ECG Wave
Option to include an ECG Wave during the test
Using Battery Performance Test
Elements
Battery Capacity Test
The Battery Capacity test is used to check whether a
battery-powered defibrillator can deliver a certain number
of discharges per minute and whether the charge time
remains adequate throughout the test. Tables 24 and 25
list the Battery Capacity test measurements and custom
parameters.