Fluke Biomedical 6000D Impulse User Manual
Page 79

Defibrillator/Transcutaneous Pacemaker Analyzer
Using Pacemaker Test Elements (Impulse 7000DP only)
65
Table 11. Pacer Parameter Test Measurements
Measurement
Unit of Measure
Description
Pacer Rate
Pulses/minute
Number of pacer pulses detected by the Analyzer during the test
Pulse Amplitude
milliamperes
Peak current detected during the test
Pulse Width
milliseconds
Pacer pulse width as measured by the Analyzer
Pacer Energy
millijoules
Pacemaker energy output
Table 12. Pacer Parameter Test Custom Parameters
Parameter Description
Input Jacks
Specifies where the pacer leads are attached on the Impulse 7000DP
Choose between pacer jacks or defib jacks.
Pacer Load
Defines load used for the test
If defib jacks are used for the test, a 50
Ω
load is used.
Brand
The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.
DUT Rate
Expected pacer rate; field user prompted to set pacemaker to this rate
DUT Amplitude
Expected pacer amplitude; field user prompted to set the pacemaker output current to this value