Defibrillator/Transcutaneous Pacemaker Analyzer
Specifications
31
Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes
.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Bi-Phasic waveform has not been approved in the United States.
Autoranged Measurement
....................................0.1 to 600 J
Accuracy
0.1 to 360 J .....................................................
±
(1 % of reading + 0.1 J)
360 to 600 J ....................................................
±
(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is
±
(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance...........................................................50
Ω
Accuracy..........................................................
±
1 %, non-inductive (<2
μ
H)
Pulse trigger level
.................................................20 V
Pulse width
Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................
±
0.1 ms
Voltage
Range..................................................................20 to 5000 V
Accuracy..............................................................
±
(1 % of reading + 2 V)
Current
Range..................................................................0.4 to 100.0 A
Accuracy..............................................................
±
(1 % of reading + 0.1 A)