Arrhythmia wave test – Fluke Biomedical 6000D Impulse User Manual
Page 86

Impulse 6000D, 7000DP
Users Manual
72
Arrhythmia Wave Test
This test is typically used to verify the shock advisory
capability of a defibrillator in response to various
arrhythmia waveforms. Table 20 lists test custom
parameters.
Note
This is a visual / audible test; the Analyzer takes
no measurements during this test.
Table 20. Arrhythmia Wave Advisory Test Custom Parameters
Parameter Description
Arrhythmia
Selects the category of arrhythmia waveform to use for the test
Arrhythmia Type
Selects the type of arrhythmia to simulate
Poly VTach Type
Provides a choice of 5 types of Poly VTach waveforms when this type of arrhythmia is selected
Mono VTach Rate
BPM value is required when Mono VTach Rate is selected.
ECG Amplitude
The voltage amplitude to use for the arrhythmia simulation
Duration
How long the simulation should last. Setting can range from 1 to 60 seconds. With checkbox
Indefinite selected, duration is ignored, and the test must be manually stopped by the field user
(Stop in TEST GUIDE toolbar).
Atrial Width, Amplitude,
and Polarity
These atrial parameters are needed for some TV Paced arrhythmia simulations.
Ventricular Width,
Amplitude, and Polarity
These ventricular parameters are needed for some TV Paced arrhythmia simulations.