Pacer sensitivity test – Fluke Biomedical 6000D Impulse User Manual
Page 81

Defibrillator/Transcutaneous Pacemaker Analyzer
Using Pacemaker Test Elements (Impulse 7000DP only)
67
Table 14. Pacer Refractory Test Custom Parameters
Parameter Description
Input Jacks
Specifies where the pacer leads are attached on the Impulse 7000
Choose between pacer jacks or defib jacks.
Pacer Load
Defines load used for the test
If defib jacks are used for the test, a 50
Ω
load is used.
Brand
The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.
DUT Rate
Expected pacer rate; field user prompted to set pacemaker to this rate
DUT Amplitude
Expected pacer amplitude; field user prompted to set the pacemaker output current to this value
Pacer Sensitivity Test
The Pacer Sensitivity test determines the threshold of
ECG amplitude required to trigger the pacemaker. Tables
15 and 16 list the Pacer Sensitivity test measurements
and custom parameters.
Table 15. Pacer Sensitivity Test Measurements
Measurement
Unit of Measure
Description
Sensitivity Threshold Amplitude
millivolts
The ECG amplitude that triggers the pacer