beautypg.com

Bronkhorst RS232 interface User Manual

Page 6

background image

BRONKHORST

®

 

Page 6   

 

 

 

 

RS232 interface   

 

 

 

9.17.027 

 
 

1.4

S

HORT FORM START

UP

 

 
 
 

All necessary settings for this module are already performed at Bronkhorst. 

To follow next steps carefully is the quickest way to get this module operational in your own system. 

 
 
 
 
 

 

 
 
 

 

In case of trouble programs like Hyper terminal (available in MS‐Windows) or FlowDDE (from 
Bronkhorst) could be very useful.
 

 
 
 
 
 

START 

Connect 
PC / PLC 

to RS232 
interface 

Check Power 

Make sure your PC or PLC is connected to the RS232 interface by means of 
the correct cable. 

 Multibus instruments need a special cable with T‐connector. 
 RS232/FLOW‐BUS interfaces need a one‐on‐one 9‐pole cable 

without crossings with male and female connector 

Cable lengths for RS232 must not exceed 10 meters. 

Make sure instrument or interface is powered (+15…24Vdc) 

Initialization procedure 

In case of RS232/FLOW‐BUS interface without micro‐switch and LED’s, first 
see that interface gets a free address on the FLOW‐BUS. Follow initialisation 
procedure described at [Initialising RS232/FLOW‐BUS interface] 

Check settings 

communication port 

Use settings [38400,n ,8 ,1 ] for your COM‐port:
Baud rate = 38K4 Baud, no parity, 8 data bits, 1 stop bit. 

Ready 

Start sending messages as described in following paragraphs. 

Start sending