Bronkhorst RS232 interface User Manual
Page 23
BRONKHORST
®
Page 23
RS232 interface
9.17.027
6
3E
Value 3E80h = 16000 = 50%
7
80
8
\r
Carriage Return
End
9
\n
Line Feed
Total String
:06030201213E80\r\n
3.10.4 Request chained parameters
The Interface sends a request for the following parameters to module at node 3:
Process 113: Serial number (3), USERTAG (6)
Process 1: Measure (0), Capacity (13), Capacity unit (31), Fluid name (17)
Request chained parameters to node 3
Nr Byte Layout
Description
Description per block
0
:
Start
1
1A
Length 26
2
03
Node 3
3
04
Command read
4
F1
11110001
C
80
Process chained (return)
Process 113 (return)
P
71
Process 113 (return)
5
EC
11101100
C
80
Parameter chained (return)
Parameter index 12 (return)
Process 113
Parameter number 3
Serial number
T
60
Parameter type ‘string’ (return)
N
0C Parameter index 12 (return)
6
71
01110001 P
71
Process 113
7
63
01100011
T
60
Parameter type ‘string’
P
03
Parameter number (FBnr.) 3 – Serial number
8
00
00000000
00
String length 00, length not defined
9
6D
01101101
C
00
Parameter not chained (return)
Parameter index 13 (return)
Process 113
Parameter number 6
USERTAG
T
60
Parameter type ‘string’ (return)
N
0D Parameter index 13 (return)
10
71
01110001 P
71
Process 113
11
66
01100110
T
60
Parameter type ‘string’
P
06
Parameter number (FBnr.) 6 – USERTAG
12
00
00000000
00
String length 00, length not defined
13
01
00000001
C
00
Parameter not chained (return)
Process 1 (return)
P
01
Process 1 (return)
14
AE
10101110
C
80
Parameter chained (return)
Parameter index 14 (return)
Process 1
Parameter number 0
Measure
T
20
Parameter type ‘integer’ (return)
N
0E
Parameter index 14 (return)
15
01
00000001 P
00
Process 1
16
20
00100000
T
20
Parameter type ‘integer’
P
00
Parameter number (FBnr.) 0 – Measure
17
CF
11001111
C
80
Process chained (return)
Parameter index 15 (return)
Process 1
Parameter number 15
Capacity
T
40
Parameter type ‘float’ (return)
N
0F
Parameter index 15 (return)
18
01
00000001 P
01
Process 1
19
4D
01001101
T
40
Parameter type ‘float’
P
0D Parameter number (FBnr.) 13 – Capacity
20
F0
11110000
C
80
Parameter chained (return)
Parameter index 16 (return)
Process 1
T
60
Parameter type ‘string’ (return)