Parameter properties table – Bronkhorst RS232 interface User Manual
Page 38
Parameter properties table
Parameter
number
(DDE)
Parameter
name
Group
0
Group
1
Group
2
Process
number
FB
nr
(par)
Var
Type
Var
Length
Min
value
Max
value
Read
Write
Poll
Secured
Highly
Secured
Default Value
Description
2011-05-18 Page 1 of 5
FlowDDE database version
V3.64
DDE str
1
Identification string
13
0
0
c
-2
Yes
Yes
No
No
No
7SN999999
identnr.+softwareversion[+serialnr.]
identstrng
2
Primary node address
1
0
1
c
0
128
Yes
Yes
No
Yes
Yes
0
primary node address: network parameter FLOW-BUS
pna
3
Secondary node address
1
0
2
c
0
128
Yes
Yes
No
Yes
Yes
0
secundary node address: network parameter FLOW-BUS
sna
4
Next node address
1
0
3
c
0
128
Yes
Yes
No
No
No
1
next node address: network parameter FLOW-BUS
nna
5
Last node address
1
0
4
c
0
128
Yes
Yes
No
No
No
32
last node address: network parameter FLOW-BUS
lna
6
Arbitrage
1
0
5
c
0
255
Yes
Yes
No
Yes
Yes
67
FLOW-BUS arbitrage setting and/or automatic optimization
arbitrage
7
Initreset
12
0
10
c
0
255
Yes
Yes
No
No
No
init and reset security key commands for network/parameter settings
initreset
8
Measure
2
0
i
-23593
41942
Yes
Yes
Yes
No
No
0
measured value (100% = 32000)
measure
9
Setpoint
2
18
1
i
0
32767
Yes
Yes
Yes
No
No
0
setpoint: wanted value (100% = 32000)
setpoint
10
Setpoint slope
18
2
i
0
30000
Yes
Yes
No
No
No
0
setpoint ramp signal 0..100 % in up to slope x 0.1 sec.
setpslope
11
Analog input
2
18
3
i
-23593
41942
Yes
No
Yes
No
No
0
analog input signal, normally used for ext. setp. (100% = 32000)
analoginp
12
Control mode
18
4
c
0
255
Yes
Yes
No
No
No
0
control mode selection for instrument or module
cntrlmode
13
Polynomial constant A
3
5
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant A (offset)
polycnst A
14
Polynomial constant B
3
6
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
1.0
polynomial constant B (span]
polycnst B
15
Polynomial constant C
3
7
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant C
polycnst C
16
Polynomial constant D
3
8
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant D
polycnst D
17
Polynomial constant E
3
9
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant E (offset) for setpoint or power value
polycnst E
18
Polynomial constant F
3
10
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
1.0
polynomial constant F (span) for setpoint or power value
polycnst F
19
Polynomial constant G
3
11
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant G for setpoint or power value
polycnst G
20
Polynomial constant H
3
12
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
0.0
polynomial constant H for setpoint or power value
polycnst H
21
Capacity
3
19
13
f
-3.40282E+38
3.40282E+38
Yes
Yes
No
Yes
No
1.0
readout value at 100% in capacity (readout) unit
capacity
22
Sensor type
3
14
c
0
255
Yes
Yes
No
Yes
No
3
sensor type information for actual reading and sensor/controller indication
sensortype
23
Capacity unit index
3
19
15
c
0
255
Yes
Yes
No
Yes
No
0
pointer to capacity (readout) unit (classic unit table]
capunit
24
Fluid number
3
16
c
0
8
Yes
Yes
No
No
No
0
fluid number: pointer to fluidset (for e.g. polynome, name and cap.)
fluidnr
25
Fluid name
3
17
c
10
Yes
Yes
No
Yes
No
AIR
name of fluid
fluidname
26
Claim node
12
18
c
0
128
Yes
Yes
No
Yes
No
0
node address of module with operation rights
claimnode
27
Modify
12
19
c
0
255
Yes
Yes
No
No
No
0
contains number(s) of changed par (0xXX par nr, 0xFF more than one par changed)
modify
28
Alarm info
4
20
c
0
255
Yes
No
Yes
No
No
0
status information of several alarms/errors in the instrument
alarminfo
29
Channel amount
17
0
12
c
1
120
Yes
Yes
No
No
No
32
amount of channels which can be operated
chanamount
30
First channel
17
0
13
c
1
120
Yes
Yes
No
No
No
1
first channel that can be operated
firstchan
31
Last channel
17
0
14
c
1
120
Yes
Yes
No
No
No
32
last channel that can be operated
lastchan
32
5
9
1
c
0
1
Yes
Yes
No
No
No
0
operation by HOST computer enable flag
hostcontrl
33
Alarm message unit type
5
10
0
c
16
Yes
No
Yes
No
No
alarm message string with unit type information
alrmmsgTA
34
Alarm message number
5
10
1
c
16
Yes
No
Yes
No
No
alarm message string with unit number information
alrmmsgnr
35
Relay status
5
10
2
c
8
Yes
No
No
No
No
status of relays/potential free contacts ('0' = not activated, '1' = activated)
relstatus
36
Actual counter value
5
0
f
0
3.40282E+38
Yes
No
Yes
No
No
0
actual value of counter
actualval
37
Signal input selection
5
1
c
8
Yes
Yes
No
No
No
signal input selection (' '=no value,'+'=pos value,'-'=neg value input)
signinpsel
38
Reset input selection
5
2
c
8
Yes
Yes
No
No
No
external reset input enable/disable ('E'=enable,' '=disable)
resinpsel
39
5
3
f
0
3.40282E+38
Yes
Yes
No
No
No
limit/batch for counter in sensor standard units
limit
40
Delay time
5
4
c
8
0
99235959
Yes
Yes
No
No
No
00000000
delay time string in days,hours,minutes,seconds
delaytime
41
Duration time
5
5
c
8
0
99235959
Yes
Yes
No
No
No
00000000
duration time string in days,hours,minutes,seconds
duratntime
42
Valve output setting
5
6
c
8
Yes
Yes
No
No
No
00000000
valve output setting ('0'=do nothing, '1'=close valve)
vlvoutset
43
Relay output setting
5
7
c
8
Yes
Yes
No
No
No
LLLLLLLL
relay output setting ('L'=low, 'H'=high, 'P'=pulse (1 sec.))
reloutset
44
Operation mode T/A
5
8
c
0
9
Yes
Yes
No
No
No
0
operation mode of T/A module
opermodeTA
45
Readout unit
5
9
c
7
Yes
No
No
No
No
ln/min
readout unit string
readunit
46
Readout factor
5
10
f
1E-10
10000000000
Yes
No
No
No
No
1
readout factor matching readout unit string
readfact
47
Reset unit
5
12
c
0
1
No
Yes
No
No
No
reset unit command (1=reset T/A unit)
resetunit
48
Valve differentiator down
6
9
f
0
3.40282E+38
Yes
Yes
No
Yes
No
0.1
valve output differentiation time constant downwards
TdValveDn
49
Valve differentiator up
6
10
f
0
3.40282E+38
Yes
Yes
No
Yes
No
0.1
valve output differentiation time constant upwards
TdValveUp
50
Sensor differentiator down
6
11
f
0
3.40282E+38
Yes
Yes
No
Yes
No
5.8
sensor signal differentiation time constant downwards
TdSensorDn
51
Sensor differentiator up
6
12
f
0
3.40282E+38
Yes
Yes
No
Yes
No
5.6
sensor signal differentiation time constant upwards
TdSensorUp
52
Cycle time
6
114
12
c
0
255
Yes
Yes
No
Yes
No
6
cycle time * 10 msec. main loop signal processing
CycleTime
53
Analog mode
10
115
3
c
0
255
Yes
Yes
No
Yes
No
0
analog mode selection for analog operation
AnalogMode
54
Reference voltage
10
116
6
i
0
65535
Yes
Yes
No
Yes
No
reference voltage output signal for analog operation
VrefOutput
55
Valve output
8
114
1
l
0
16777215
Yes
Yes
Yes
No
No
0
valve output signal (24-bit number in range 0..14.3Vdc/0..23.3Vdc)
ValveOut
56
Dynamic display factor
6
117
1
f
0
1
Yes
Yes
No
Yes
No
0,001
dynamic display factor for display filter (0=max, 1=min goes with par 57)
DynDispFct
57
Static display factor
6
117
2
f
0
1
Yes
Yes
No
Yes
No
0,000001
static display factor for display filter (0=max, 1=min goes with par 56)
StaDispFct
58
Calibration mode
7
115
1
c
0
255
Yes
Yes
No
Yes
No
0
calibration mode selection (not active until cntrlmode has been set to value 9)
CalMode
59
Valve offset
8
116
7
i
-32767
65535
Yes
Yes
No
Yes
No
61000
valve offset: amount of DAC steps within 1 potmeter step
ValveOffst
60
Monitor mode
2
115
2
c
0
255
Yes
Yes
No
Yes
No
7
monitor: output signal (measure) selection for bus and analog output
Monitor
61
Alarm register1
4
114
2
c
8
Yes
Yes
No
Yes
No
alarm register containing warning flags
AlarmReg1
62
Alarm register2
4
114
3
c
8
Yes
Yes
No
Yes
No
alarm register containing critical error flags
AlarmReg2
63
9
116
1
l
0
16777215
Yes
Yes
No
Yes
No
210A7D
calibration register zero scale input 1 ADC
CalRegZS1
64
9
116
2
l
0
16777215
Yes
Yes
No
Yes
No
52A513
calibration register full scale input 1 ADC
CalRegFS1
65
9
116
3
l
0
16777215
Yes
Yes
No
Yes
No
210A7D
calibration register zero scale input 2 ADC
CalRegZS2