Bronkhorst RS232 interface User Manual
Page 35
Parameter
number
(DDE)
Parameter
name
Filter
Value
Description
2011-05-18 Page 3 of 5
FlowDDE database version
V3.64
86
IO status
&H40
analog jumper set (read only)
1
86
IO status
&H80
micro switch pressed (read only)
1
106
Pressure sensor type
delta-P 0..5" W.C.
0
106
Pressure sensor type
delta-P 0...10" W.C.
1
106
Pressure sensor type
absolute pressure 800-1200 mbar
2
106
Pressure sensor type
absolute pressure 800-1100 mbar
3
106
Pressure sensor type
delta-P -5...0 "W.C.
4
106
Pressure sensor type
delta-P -10...0 "W.C.
5
106
Pressure sensor type
delta-P -10...+10 "W.C.
6
106
Pressure sensor type
delta-P 0...1 PSI
7
106
Pressure sensor type
delta-P -1...0 PSI
8
106
Pressure sensor type
absolute pressure 0-10 bar
10
114
Reset
no reset
0
114
Reset
reset counter value (no mode change) or common reset
1
114
Reset
reset alarm
2
114
Reset
restart batch counter
3
114
Reset
reset counter value (counter off)
4
114
Reset
Reset module (soft reset)
5
118
Alarm mode
off
0
118
Alarm mode
alarm on absolute limits
1
118
Alarm mode
alarm on limits related to setpoint (response alarm)
2
118
Alarm mode
alarm when instrument powers-up (eg. after power-down)
3
119
Alarm output mode
no relais activity at alarm
0
119
Alarm output mode
relais pulses until reset
1
119
Alarm output mode
relais activated until reset
2
120
Alarm setpoint mode
no setpoint change at alarm
0
120
Alarm setpoint mode
new/safe setpoint at alarm enabled (set at par 121)
1
125
Counter output mode
no relais activity at batch limit
0
125
Counter output mode
relais pulses after reaching batch limit until reset
1
125
Counter output mode
relais activated after reaching batch limit until reset
2
126
Counter setpoint mode
setpoint change at batch limit disabled
0
126
Counter setpoint mode
setpoint change at batch limit enabled
1
130
Counter mode
off
0
130
Counter mode
counting upwards continuously
1
130
Counter mode
counting up to limit (batchcounter)
2
147
Range select
calibration ready/stop
0
147
Range select
run calibration until stopsensor 1/select range 1
1
147
Range select
run calibration until stopsensor 2/select range 2
2
147
Range select
run calibration until stopsensor 3/select range 3
3
147
Range select
run calibration until stopsensor 4/select range 4
4
147
Range select
run calibration and select range 5
5
147
Range select
run calibration with automatic range selection
9
147
Range select
run until stopsensor 1 until 3 values between limit
19
147
Range select
run until stopsensor 2 until 3 values between limit
29
147
Range select
run until stopsensor 3 until 3 values between limit
39
147
Range select
run until stopsensor 4 until 3 values between limit
49
147
Range select
run and select range 5 until 3 values between limit
59
147
Range select
run with auto-select + 3 values between limit
99
156
Reset alarm enable
no reset possible
0
156
Reset alarm enable
reset: keyboard
1
156
Reset alarm enable
reset: external
2
156
Reset alarm enable
reset: keyboard or external
3
156
Reset alarm enable
reset: FLOW-BUS
4
156
Reset alarm enable
reset: FLOW-BUS or keyboard
5
156
Reset alarm enable
reset: FLOW-BUS or external
6
156
Reset alarm enable
reset: FLOW-BUS or keyboard or external
7
156
Reset alarm enable
reset: automatic
8
156
Reset alarm enable
reset: automatic or keyboard
9
156
Reset alarm enable
reset: automatic or external
10
156
Reset alarm enable
reset: automatic or keyboard or external
11
156
Reset alarm enable
reset: automatic or FLOW-BUS
12
156
Reset alarm enable
reset: automatic or FLOW-BUS or keyboard
13
156
Reset alarm enable
reset: automatic or FLOW-BUS or external
14
156
Reset alarm enable
reset: automatic or FLOW-BUS or keyboard or external
15
157
Reset counter enable
no reset possible
0
157
Reset counter enable
reset: keyboard
1
157
Reset counter enable
reset: external
2
157
Reset counter enable
reset: keyboard or external
3
157
Reset counter enable
reset: FLOW-BUS
4
157
Reset counter enable
reset: FLOW-BUS or keyboard
5
157
Reset counter enable
reset: FLOW-BUS or external
6
157
Reset counter enable
reset: FLOW-BUS or keyboard or external
7
157
Reset counter enable
reset: automatic
8
157
Reset counter enable
reset: automatic or keyboard
9
157
Reset counter enable
reset: automatic or external
10
157
Reset counter enable
reset: automatic or keyboard or external
11
157
Reset counter enable
reset: automatic or FLOW-BUS
12
157
Reset counter enable
reset: automatic or FLOW-BUS or keyboard
13
157
Reset counter enable
reset: automatic or FLOW-BUS or external
14
157
Reset counter enable
reset: automatic or FLOW-BUS or keyboard or external
15
166
Controller features
&H01
valve in normal position after startup
0
166
Controller features
&H01
valve in safe position after startup
1
166
Controller features
&H02
open from zero with PID output to valve
0
166
Controller features
&H02
open from zero with ramp output to valve
1