Bronkhorst RS232 interface User Manual
Page 34
Parameter
number
(DDE)
Parameter
name
Filter
Value
Description
2011-05-18 Page 2 of 5
FlowDDE database version
V3.64
58
Calibration mode
adjust Vref output by connecting it to analog in
10
58
Calibration mode
adjust analog out by connecting it to analog in
11
58
Calibration mode
adjust valveoutput by connecting it to analog in
12
58
Calibration mode
dmfc
13
58
Calibration mode
dmfc
14
58
Calibration mode
analog output = 0 %
15
58
Calibration mode
analog output = 100 %
16
58
Calibration mode
analog output = 50 %
17
58
Calibration mode
factory zero
18
58
Calibration mode
sensor differentiator (setpoint steps are needed!)
19
58
Calibration mode
automatic sensor configuration
20
58
Calibration mode
sensor temperature calibration
21
58
Calibration mode
customer zero (no control mode 9 needed)
22
58
Calibration mode
Error mode (result of previous cal mode)
255
60
Monitor mode
(filtered) setpoint
0
60
Monitor mode
controller error input signal / raw sensor signal
1
60
Monitor mode
controller output signal to valve
2
60
Monitor mode
sensor signal slow
3
60
Monitor mode
sensor signal slow filtered
4
60
Monitor mode
linearization output
5
60
Monitor mode
differentiator output
6
60
Monitor mode
differentiator output filtered
7
60
Monitor mode
normal sensor signal (Output)
8
60
Monitor mode
analog input signal
9
60
Monitor mode
power supply voltage
10
60
Monitor mode
mass flow in display unit (normally ln/min)
11
60
Monitor mode
volume flow in l/min
12
60
Monitor mode
temperature in °C
13
60
Monitor mode
pressure absolute in mbara
14
60
Monitor mode
time in msec/frequency in Hz.
15
60
Monitor mode
calibrated volume at actual sensor in ml
16
60
Monitor mode
delta-P pressure in mbarg
17
60
Monitor mode
atmospheric (barometer) pressure in mbara
18
60
Monitor mode
mass flow in kg/min
19
61
Alarm register1
&H8000000000000000
No diagnostics available in warning register
0
61
Alarm register1
&H8000000000000000
Diagnostics available in warning register
1
62
Alarm register2
&H8000000000000000
No diagnostics available in error register
0
62
Alarm register2
&H8000000000000000
Diagnostics available in error register
1
67
ADC control register
&H001000
ADC bipolar mode
0
67
ADC control register
&H001000
ADC unipolar mode
1
67
ADC control register
&H1000000
Disable zero measure threshold
0
67
ADC control register
&H1000000
Enable zero measure threshold
1
67
ADC control register
&H1C0000
ADC gain = 1x
0
67
ADC control register
&H1C0000
ADC gain = 2x
1
67
ADC control register
&H1C0000
ADC gain = 4x
2
67
ADC control register
&H1C0000
ADC gain = 8x
3
67
ADC control register
&H1C0000
ADC gain = 16x
4
67
ADC control register
&H1C0000
ADC gain = 32x
5
67
ADC control register
&H1C0000
ADC gain = 64x
6
67
ADC control register
&H1C0000
ADC gain = 128x
7
69
disable
0
69
enable
1
70
Test mode
idle; no action
0
70
Test mode
uProcessor
1
70
Test mode
IO
2
70
Test mode
RAM
3
70
Test mode
FRAM
4
70
Test mode
ADC
5
70
Test mode
DAC
6
70
Test mode
sensor
7
70
Test mode
valve drive circuit
8
70
Test mode
Vref
9
70
Test mode
FLOW-BUS
10
70
Test mode
calibration
11
70
Test mode
keyboard
12
71
AD channel 1
1
71
AD channel 2
2
79
Tuning mode
idle; no action
0
79
Tuning mode
sensor
1
79
Tuning mode
valve
2
79
Tuning mode
Fuzzy controller normal operation
3
79
Tuning mode
Fuzzy controller open at zero
4
79
Tuning mode
PID controller
5
80
Valve default
normally closed
0
80
Valve default
normally opened
1
80
Valve default
normally closed inverse controlled
2
80
Valve default
normally opened inverse controlled
3
80
Valve default
remain position
4
86
IO status
&H01
read diagnostic jumper (no diagnostics, read/write)
1
86
IO status
&H02
not used
1
86
IO status
&H04
read analog jumper (use cntrlmode, read/write)
1
86
IO status
&H08
read micro switch (read/write)
1
86
IO status
&H10
diagnostic jumper set (read only)
1
86
IO status
&H20
initialization jumper set (read only)
1