Bronkhorst RS232 interface User Manual

Page 41

background image

Parameter

number

(DDE)

Parameter

name

Group

0

Group

1

Group

2

Process

number

FB
nr

(par)

Var

Type

Var

Length

Min

value

Max

value

Read

Write

Poll

Secured

Highly

Secured

Default Value

Description

2011-05-18 Page 4 of 5

FlowDDE database version

V3.64

DDE str

206

fSetpoint

2

18

33

3

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

No

No

0

setpoint: wanted value for direct reading (in capunits, max.= capacity)

fSetpoint

207

Mass

20

33

23

f

-3.40282E+38

3.40282E+38

Yes

No

Yes

No

No

0

Mass in g

Mass

208

Manufacturer status register

4

119

1

c

8

Yes

No

No

No

No

Manufacturer Status register (64 diagnostic bits)

Mstatus

209

Manufacturer warning register

4

119

2

c

8

Yes

No

No

No

No

Manufacturer Warning register (64 diagnostic bits)

Mwarning

210

Manufacturer error register

4

119

3

c

8

Yes

No

No

No

No

Manufacturer Error register (64 diagnostic bits)

Merror

211

Diagnostic history string

4

119

4

c

-2

Yes

Yes

No

No

No

Diagnostic history string (contains history of diag codes)

DiagHist

212

Diagnostic mode

4

119

5

c

0

255

Yes

Yes

No

Yes

No

0

Diagnostic mode (0 = diagnostics off, 1 = diagnostics on)

DiagMode

213

Manufacturer status enable

4

119

6

c

0

255

No

Yes

No

No

No

0

Manufacturer Status enable (0-127 or 254 = disable all, 255 = enable all)

MStatEnabl

214

Analog output zero adjust

10

116

21

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Analog measure output, zero adjust

AnOutZA

215

Analog output span adjust

10

116

22

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Analog measure output, span adjust

AnOutSA

216

Analog input zero adjust

10

116

23

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Analog setpoint input, zero adjust

AnInZA

217

Analog input span adjust

10

116

24

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Analog setpoint input, span adjust

AnInSA

218

Sensor input zero adjust

10

116

25

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Sensor input, zero adjust

SensInZA

219

Sensor input span adjust

10

116

26

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Sensor input, span adjust

SensInSA

220

Temperature input zero adjust

10

116

27

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Sensor temperature input, zero adjust

TempInZA

221

Temperature input span adjust

10

116

28

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Sensor temperature input, span adjust

TempInSA

222

Adaptive smoothing factor

6

117

5

f

0

1

Yes

Yes

No

Yes

No

1.0

Sensor input filter adapt setting

ExpSmooAd

223

Slope setpoint step

18

33

24

i

0

32000

Yes

Yes

No

Yes

No

32000

Slope setpoint step. Setpoint step for the given slopetime

SlopeSetp

224

Filter length

6

117

6

i

0

255

Yes

Yes

No

Yes

No

1

Number of samples for average filter

FilterLen

225

Absolute accuracy

2

33

25

f

-3.40282E+38

3.40282E+38

Yes

No

No

No

No

0

Actual accuracy in current unit

fAccuracy

226

Lookup table index

3

33

26

c

0

255

Yes

Yes

No

No

No

0

Lookup table for linearisation index (x and y direction)

LookI

227

Lookup table X

3

33

27

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for linearisation x

LookX

228

Lookup table Y

3

33

28

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for linearisation y

LookY

229

Lookup table temperature index

3

33

29

c

0

255

Yes

Yes

No

No

No

0

Lookup table for linearisation at certain temperature index (z direction)

LookTempI

230

Lookup table temperature

3

33

30

f

-273.15

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for linearisation at certain temperature (z)

LookTemp

231

Valve maximum

8

114

25

f

0

24

Yes

Yes

No

Yes

No

0.2

Maximum current/voltage for valve

ValveMax

232

Valve mode

8

114

26

c

0

255

Yes

Yes

No

Yes

No

1

Valve output mode selection

ValveMode

233

Valve open correction

8

114

27

f

0

1.5

Yes

Yes

No

Yes

No

0.96

Valve open current/voltage correction (example: 0.96, Open = ValveOpen * 0.96)

VlvOpenCor

234

Valve zero hold

8

114

28

f

0

1

Yes

Yes

No

Yes

No

0

Valve hold current/voltage at %0 setp (example: 0.8, Hold = ValveOpen * 0.8)

VlvZeroHld

235

Valve slope

8

114

29

f

0

50

Yes

Yes

No

Yes

No

0.009

Valve slope time (seconds)

ValveSlope

236

IFI data

23

0

21

c

-2

Yes

Yes

No

No

No

0

IFI data dump protocol communication string

IFIData

237

Range used

20

19

115

20

c

-2

0

99

Yes

No

Yes

No

No

0

Piston Prover information about used sensors

RangeUsed

238

Fluidset properties

3

33

31

c

0

255

Yes

Yes

No

Yes

Yes

0

Fluidset properties

FldSetProp

239

Lookup table unit type index

3

33

12

c

0

255

Yes

Yes

No

Yes

Yes

0

Lookup table unit type

LUnitType

240

Lookup table unit type

3

33

13

c

20

Yes

Yes

No

Yes

Yes

Lookup table unit type name

LUnTypName

241

Lookup table unit index

3

33

16

c

0

255

Yes

Yes

No

Yes

Yes

0

Lookup table unit (unit LUTy)

LUnit

242

Lookup table unit

3

33

17

c

7

Yes

Yes

No

Yes

Yes

kg/s

Lookup table unit name

LUnitName

243

Capacity unit type index

3

29

c

0

255

Yes

Yes

No

Yes

No

0

Capacity (readout) unit type

CUnitType

244

Capacity unit type

3

30

c

20

Yes

Yes

No

Yes

No

Capacity (readout) unit type name

CUnTypName

245

Capacity unit type temperature

3

33

10

f

-273.15

3.40282E+38

Yes

Yes

No

Yes

No

0

Capacity (readout) unit type temperature (°C)

CUnTypTemp

246

Capacity unit pressure

3

33

11

f

0

3.40282E+38

Yes

Yes

No

Yes

No

1

Capacity (readout) unit type pressure (bar (a))

CUnTypPres

247

Capacity minimum

3

27

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Minimum capacity in output capacity units

CapMin

248

Capacity maximum

3

28

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Maximum capacity in output capacity units

CapMax

249

Formula type

3

113

17

i

0

65535

Yes

Yes

No

Yes

No

0

Formula type needed for conversion

FormulaTyp

250

Heat capacity

3

113

18

f

0

3.40282E+38

Yes

Yes

No

Yes

No

0

Heat capacity (Cp) (sensor conditions)

HeatCap

251

Thermal conductivity

3

113

20

f

0

3.40282E+38

Yes

Yes

No

Yes

No

0

Thermal conductivity (sensor conditions)

ThermCond

252

Viscosity

3

113

21

f

0

3.40282E+38

Yes

Yes

No

Yes

No

0

Dynamic viscosity (fluid conditions)

Viscosity

253

Standard flow

3

113

22

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

1

Standard mass flow in ln/min (20°C, 1.01325 bar (a) ) air or g/h H2O equivalent

NormMasFlw

254

Controller speed

8

3

114

30

f

0

3.40282E+38

Yes

Yes

No

Yes

No

1

Controller speed factor (gain)

Kspeed

255

Sensor code

13

113

23

i

0

65535

Yes

Yes

No

Yes

Yes

0

Sensor code

SensorCode

256

Sensor configuration code

13

113

24

c

0

255

Yes

Yes

No

Yes

Yes

0

Sensor configuration code

SensorRevC

257

Restriction code

13

113

25

i

0

65535

Yes

Yes

No

Yes

Yes

0

Restriction code

RestrCode

258

Restriction configurator code

13

113

26

c

0

255

Yes

Yes

No

Yes

Yes

0

Restriction configuration code

RestrRevC

259

Restriction NxP

13

113

27

l

0

2147483648

Yes

Yes

No

Yes

Yes

0

Restriction NxP (proportional to air equivalent capacity of LFE)

RestrNxP

260

Seals information

13

113

28

c

16

0

255

Yes

Yes

No

Yes

Yes

V,V

Seals information (1st byte = other, 2nd = plunger seal)

Seals

261

Valve code

13

113

29

i

0

65535

Yes

Yes

No

Yes

Yes

0

Valve code

ValveCode

262

Valve configuration code

13

113

30

c

0

255

Yes

Yes

No

Yes

Yes

0

Valve configuration code

ValveRevC

263

Instrument properties

13

113

31

l

0

2147483648

Yes

Yes

No

Yes

Yes

0

Instrument properties

InstrProp

264

Lookup table frequency index

3

116

10

c

0

1

Yes

Yes

No

Yes

No

0

Lookup table for frequency index

LookFreqI

265

Lookup table frequency frequency

3

116

11

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for frequency frequency

LFFreq

266

Lookup table frequency temperature

3

116

12

f

-273.15

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for frequency temperature

LFTemp

267

Lookup table frequency density

3

116

13

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for frequency density

LFDensity

268

Lookup table frequency span adjust

3

116

14

f

-3.40282E+38

3.40282E+38

Yes

Yes

No

Yes

No

0

Lookup table for frequency span adjust

LFSpanAdj

269

Capacity unit index (ext)

3

19

65

15

c

0

255

Yes

Yes

No

Yes

No

0

Capacity (readout) unit index (extended unit table)

CUnit

270

Density actual

3

116

15

f

-3.40282E+38

3.40282E+38

Yes

No

Yes

No

No

0

Actual density, measured by instrument

DensityAct

271

Measured restriction

13

116

18

f

0

3.40282E+38

Yes

Yes

No

Yes

No

0

Measured restriction

RestrMeas

272

Temperature potmeter

3

116

8

c

0

255

Yes

Yes

No

Yes

No

0

Potmeter for sensor temperature compensation

TempPotm

273

Temperature potmeter gain

3

116

9

c

0

255

Yes

Yes

No

Yes

No

0

Gain for sensor temperature compensation

TempGain

274

Counter controller overrun correction

16

104

10

f

0

3.40282E+38

Yes

Yes

No

No

No

1

Counter controller overrun correction

CntrCConv

275

Counter controller gain

16

104

11

f

0

3.40282E+38

Yes

Yes

No

No

No

1

Counter controller gain

CntrCGain