3 capacities, 4 pressure rating, 5 piping – Bronkhorst Vapor Delivery Module User Manual
Page 9: 6 electrical connections, 1 mains, Capacities, Pressure rating, Piping, Electrical connections, Mains

Bronkhorst®
Vapor Delivery Module
9
9.17.079
3.4.3
Capacities
Figure 1: Capacity SW-100 Vaporizer Figure 2: Capacity SW-200 Vaporizer
3.4.4
Pressure rating
• Gas supply = max. 10 bar(g)
• Liquid supply = max. 10 bar(g)
• Vapor pressure = atmospheric
3.5
Piping
For optimal performance of the system it is recommend to have at least 10 pipe diameters distance between the angle and the
equipment inlets. The purge is rarely used in production environments, except for installation.
Be sure that piping is absolutely clean!
CAUTION - DO NOT install small diameter piping on high flow rates.
CAUTION - DO NOT mount abrupt angles directly to the inlet or outlet, especially for high flow rates.
3.6
Electrical connections
3.6.1
Mains
Model
Power
supply
(Vac)
Maximum power
(included tracing
§3.6.2)
Fuse primary heater
Fuse secondary
heater
Fuse main
SW-10n-n-50-a-nn-a
110 … 120
250VA 50-60 Hz
T2A I2t >30A
Littelfuse 0213002.MXP*
T4A
T2A
SW-10n-n-51-a-nn-a
110 … 120
550VA 50-60 Hz
T2A I2t >30A
Littelfuse 0213002.MXP*
T4A
T2A
SW-20n-n-50-a-nn-a
110 … 120
250VA 50-60 Hz
T2A I2t >30A
Littelfuse 0213002.MXP*
T4A
T2A
SW-20n-n-51-a-nn-a
110 … 120
550VA 50-60 Hz
T2A I2t >30A
Littelfuse 0213002.MXP*
T4A
T2A
SW-10n-n-30-a-nn-a
220 … 240
250VA 50-60 Hz
T1A I2t >10A
Littelfuse 0213001.MXP*
T4A
T1A
SW-10n-n-31-a-nn-a
220 … 240
550VA 50-60 Hz
T1A I2t >10A
Littelfuse 0213001.MXP*
T4A
T1A
SW-20n-n-30-a-nn-a
220 … 240
250VA 50-60 Hz
T1A I2t >10A
Littelfuse 0213001.MXP*
T4A
T1A
SW-20n-n-31-a-nn-a
220 … 240
550VA 50-60 Hz
T1A I2t >10A
Littelfuse 0213001.MXP*
T4A
T1A
* Littelfuse or equivalent