2 starting up, 1 check properties, 2 rated pressure test inspection – Bronkhorst Vapor Delivery Module User Manual

Page 7: 3 install module, 4 model key, Starting up, Check properties, Rated pressure test inspection, Install module, Model key

background image

Bronkhorst®

Vapor Delivery Module

7

9.17.079

2

Starting up

2.1

Check properties

Before installing a Vapor Delivery Module it is important to read the attached labels

and check:

Model key
Pressure rate
Temperature (ambient)
Power supply
Power

2.2

Rated pressure test inspection

Each Vapor Delivery Module is pressure tested to at least 1.5 times the working pressure of the process conditions stipulated by the

customer, with a minimum of 8 bar. Each Module is helium leak tested to at least 2•10

-9

[mbar l/s] Helium outboard.

The tested pressure is stated on the module with a RED COLOURED sticker.

Before installation, make sure that the test pressure is in accordance with

normal safety factors for your application.

If there is no Pressure Testing Sticker on the module or if the test pressure is

not complying with your pressure requirements, the equipment should not

be mounted in the process line and be returned to the factory.

2.3

Install module

See chapter 4 for detail installation instructions.

Before switching on power, please check if all external electrical and communication (if necessary) connections are properly

connected.

2.4

Model key