beautypg.com

3 technical specification, 1 physical dimensions, 2 environmental conditions – Bronkhorst Vapor Delivery Module User Manual

Page 8: 3 ingress protection class, 4 fluidic connections, 1 general, 2 flow ranges of used components, Technical specification, Physical dimensions, Environmental conditions

background image

Bronkhorst®

Vapor Delivery Module

9.17.079

8

3

Technical specification

3.1

Physical dimensions

i

www

For overall dimensions please consult supplied dimensional drawing 7.05.901. This documents is available on:

http://www.bronkhorst.com/en/downloads or can be obtained by contacting our local sales & service

representatives

3.2

Environmental conditions

• Ambient temperature

: 5°C - 40°C.

• Relative humidity

: Maximum relative humidity 80%, for temperatures up to 31°C decreasing linearly

to 50% relative humidity at 40°C.

• Height altitude

: Up to 2000 m.

3.3

Ingress protection class

IP20, according IEC standard 60529.

3.4

Fluidic connections

3.4.1

General

The Vapor Delivery Module gas and liquid supply are equipped with compression type or face-seal-fittings. Normally these fittings

are BSP parallel threads types which have to be used in combination with elastomeric O-rings to seal to the Vapor Delivery Module.

Occasionally fittings are orbital welded to the VDM-Series.

For leak tight installation of compression type fittings be sure that the tube is inserted to the shoulder in the fitting body and that

no dirt or dust is present on tube, ferrules or fittings.

For exact information on the fitting, please contact Swagelok for further instructions or documentation.

Always check your system for leaks, before applying fluid pressure. Especially if toxic, explosive or other dangerous

fluids are used.

3.4.2

Flow ranges of used components

Model

Heater

model

Thermal

power

Max. capacity

1

Max temp.

SW-10n

W-100A

10 W

1...30 g/h liquid

0.1...4 l

n

/min gas

200°C

SW-20n

W-200A

100 W

10...120 g/h liquid

0.2...10 l

n

/min gas

200°C

1

Depends on liquid and process pressure; table based on water at 1 atmosphere.

For other liquids or pressure conditions please consult the manufacturer or your local sales representative.