4.6.2

Startup with using vacuum

4.6.2.1 To fill the system

Procedure:

Actions:

A1

Evacuate the lines “Vacuum”

Start pumps

A2

Evacuate all lines

Close: V1, V2, V3, V4, and V6
Make sure that “waste” and “process” are vacuum
Connect inert gas on carrier gas inle.
Open : V5 and V8
Give setpoint to gas and liquid control (the control valves

of the LFC and MFC open)
Pump for a long time
Remark: if the system is not completely dry follow

procedure of E3, E4, (E5)

A3

Purge line between inert gas source and valve V1

Open valve V1 briefly for some seconds.

A4

Purge line between process gas source and gas MFC

Give setpoint = 0 to LFM (mix valve closes)
Open valve V3 briefly for some seconds

A5

Pump down liquid supply

Open V2 for some time
Close V2

A6

Setup carrier gas

Close V8
Open V3

A7

Fill the system with liquid

Close V1
Open B2 briefly; allow mixing valve VDM to fill
Close V5

Make sure that the carrier gas is always flowing when the

liquid flow is on. Preferably, let always gas flow in the

carrier gas line when the system is filled with liquid.

A8

Optional: degassing mixing valve

Open V5, a needle valve will be best, to avoid too much

liquid be pumped away

A9

The system is now ready for use

4.6.2.2 To start the process

Procedure:

Actions:

B1

Set the heater at right temperature

Give setpoint to the heater, wait until temperature is

reached

B2

Start gas flow

Give setpoint to MFC

Make sure that the carrier gas is always flowing when the

liquid flow is on. Preferably, let always gas flow in the

carrier gas line when the system is filled with liquid.

B3

Start liquid flow

Give setpoint LFC, wait until stabilization

Close supply pressure of the liquid vessel during the process.

The pressure controller of the inert gas supply can influence

the stability of the liquid flow controller.

B4

Connect process

Open V7, close V6