beautypg.com

3 operation, 1 general, 2 power and warm-up – Bronkhorst Mass Flow User Manual

Page 20: 3 zeroing, 4 start-up, 5 operating conditions

background image

BRONKHORST HIGH-TECH B.V.

3 Operation

3.1 General

The Bronkhorst High-Tech instruments are designed in such a way that they will meet user process

requirements in the best possible way.

Flow / pressure meters and controllers are powered from a dc power source. When providing your own

power supply be sure that voltage and current rating are according to the specifications of the instrument(s)

and furthermore that the source is capable of delivering enough energy to the instrument(s).

Cable wire diameters should be sufficient to carry the supply current and voltage losses must be kept as low

as possible. When in doubt: consult factory.

3.2 Power and warm-up

Before switching on power check if all connections have been made according to the hook-up diagram,

which belongs to the instrument.

It is recommended to turn on power before applying pressure on the instrument and to switch off power after

removing pressure.

Check fluid connections and make sure there is no leakage. If needed purge the system with a proper fluid.

For a gas instrument only purging with gases is allowed. Liquid instruments may be purged with either a gas

or a liquid, whatever is needed for the purpose.

Turn on power and allow at least 30 minutes to warm up and stabilize. In cases where no electronics are

involved (valves only) warming up is not needed.

During warm-up period, fluid pressure may either be on or off.

3.3 Zeroing

In general the zero point of each instrument is factory adjusted. If so required the zero point of the instrument

may be re-adjusted.

Zero points must be adjusted to read approx. 0.2%. The circuit will cut off signals below zero to prevent

negative zero drift. When adjustment is required, first make sure that there is a positive deviation, and then

slowly adjust the readout to read approx. 0.2%. To be sure that there is no accidental flow or pressure level

other than zero, adjustment is done best when the instrument is not part of a system.

3.4 Start-up

Turn on fluid supply gently. Avoid pressure shocks, and bring the instrument gradually up to the level of the

actual operating conditions. Also switch off fluid supply gently. In case of liquid control be sure to remove all

trapped gas bubbles from the system. The purge connection on top of the control valve can be used for this

purpose.

3.5 Operating conditions

Each instrument has been calibrated and adjusted for customer process conditions.

Controllers or valves may not operate correctly, if process conditions vary too much, because of the

restriction of the orifice in the valve.

For flowmeters performance and accuracy may be affected tremendously if physical fluid properties such as

heat capacity and viscosity change due to changing process conditions.

page 20

9.17.001