7 electronics – Bronkhorst Mass Flow User Manual
Page 11

BRONKHORST HIGH-TECH B.V.
1.7 Electronics
Each electronic housing is designed to provide RFI and EMI protection.
The p.c.boards designed by BRONKHORST HIGH-TECH B.V. are mainly provided with surface mounted
devices (SMD).
Each electronic p.c.board is set for one of the following output (and corresponding input) signals:
Signal
output (sensor)
input (setpoint)
code
signal
signal
A
0…5 Vdc
0…5 Vdc
B
0…10 Vdc
0…10 Vdc
F
0…20 mA (sourcing)
0…20 mA (sinking)
G
4…20 mA (sourcing)
4…20 mA (sinking)
K
0…5 Vdc (cable compensation)
n.a.
For meters only
L
0…10 Vdc (cable compensation)
n.a.
For meters only
mA
- +
instrument
output
0V common
power
supply
output
Sinking
Sourcing
Current output signals
I
mA
+
-
0V common
instrument
I
For meters only the output signal is available.
9.17.001
page 11