14 electromagnetic compatibility – Bronkhorst Mass Flow User Manual
Page 18
BRONKHORST HIGH-TECH B.V.
2.14 Electromagnetic compatibility
2.14.1 Conditions for compliance with EMC requirements
All instruments described in this manual carry the CE-mark.
Therefore they have to comply with the EMC requirements as are valid for these instruments.
However compliance with the EMC requirements is not possible without the use of proper cables and
connector/gland assemblies.
For good results Bronkhorst High-Tech B.V. can provide standard cables. Otherwise follow the guidelines as
stated below.
1. D-Connector assembly
Fold the shield of the cable back over the cable (the shield must be around the cable).
Wind a copper tape around the shield
Solder a black wire on the tape and
connect to pin 9 of connector
copper tape
shielded cable
e.g. LAPP LiYCY
other wires
D-connector housing
metalized
connector
8 mm
20 mm
black wire
(shield)
2. Cable gland assembly
Fold the shield of the cable back over the cable (shield must be around the cable).
shielded cable
e.g. LAPP LiYCY
shield
metal cable gland
e.g. HUMMEL HSK-M-EMV
shield
15 mm
35 mm
Mount the metal PG cable gland as shown in the drawing above
page 18