1.8.2 Liquid conversion factors

The general formula for determining the relationship between signal and mass flow reads:

V

k c

signal

p

m

= ⋅

⋅ Φ

in which:

V

signal

= output signal

k

= calibration constant

c

p

= heat capacity at constant pressure of the fluid

Φ

m

= mass flow

A conversion factor must be used if the liquid flow meter is not used on the calibrated liquid.

This conversion factor reads:

1

2

m

m

Cf Φ

=

Φ

Cf c

c

p1

p2

=

in which:

c

p1

= heat capacity of the calibration liquid

c

p2

= heat capacity of the new liquid

For application of this formula consult Bronkhorst High-Tech B.V.

1.8.3 Software for conversion factor calculation

Bronkhorst High-Tech B.V. gathered the physical properties of over 600 fluids in a database called

FLUIDAT.

Application software, such as FLUIDAT on the Net (FOTN), enable the user to calculate accurate

conversion factors, not only at 20°C/1 atm (as shown in the conversion table, App.1) but at any

temperature/pressure combination.

Apply to your distributor for more details of this software.