AMETEK SGA Series User Manual
Page 25

Sorensen SGA Series
Installation
M550129-01 Rev AG
2-5
WARNING!
A minimum two-person lift is required for the 3U SGA Series power
supply, which weighs up to 110 lb (50 kg) depending on the model.
Refer to Figure 2-1 for 3U rack mount assembly drawing for the following
instructions:
1. Install the slide sections, 1C , on both sides of the power supply chassis with
screws (three on left side and four on right side). Ensure that the latch spring
orientation is as shown in Note 1 .
2. Install the brackets, 2A and 2B , to the stationary slide sections, 1A , with
the hardware provided by the slide supplier as shown in Note 3 . Ensure that
stopper orientation of slide sections, 1B , is as shown in Note 2 . Adjust the
location of the mounting brackets as required for the particular type of cabinet
vertical rails utilized.
3. Mount the stationary slide sections, 1A , (with brackets already installed) into
the cabinet using appropriate hardware (e.g. bar nuts, cage nuts, clip nuts),
while ensuring that they are level front to back and left to right of the cabinet
rails.
4. Insert power supply chassis with slide sections, 1C , into slide sections, 1B .
5. Secure the front panel of the power supply chassis to the cabinet rack rails
using the screws provided in the ship kit.