Casella CEL Microdust 880 IS real time dust monitor User Manual
Page 37

Microdust 880nm & WinDust Application Software
Handbook HB3255-03
Page 37 of 51
If no data is sent or received by the modem during the specified time period then the
communications session will be terminated.
The modem connected to the Microdust should be configured as follows:-
AT&F8
factory defaults used by AT&F
AT&F
Load factory defaults
AT&D0
Ignore DTR status
ATE0
Turn echo off
ATQ1
Disable result codes
AT&W0
Save current settings and load them at power up.
The office based modem and P.C software should be configured for 'hardware' flow control
(AT&E4).
Above settings are for a Multitec MT1432MK modem. The required settings may vary for other