Model 4857 process controller specifications – Parr Instrument 4857 User Manual

Page 9

background image

4857 Reactor Controllers

- 9 -

M

ODEL

4857

P

ROCESS

C

ONTROLLER

S

PECIFICATIONS

Temperature Control Parameters


To facilitate start up with reasonably
good PID control parameters for heating
and cooling, all 4857 Process
Controllers are programmed at the
factory with the temperature control
settings shown on the right.


Output 1 on these controllers drives the
heater. Output 2 can be used to
activate a cooling system. Control
parameters can be set independently for
each of these outputs.

Module 1: Temperature Control

Cycle time

2 seconds

Proportional Band

25

°C

Integral Time

5 min

Derivative Time

25 sec


Input thermocouple

Type J

Operating range

0 to 800

°C

Display resolution

.1

°C

Setpoint resolution

.1

°C

System accuracy

+/- 0.1% FS
+/- 1°C typical

Control action

Three term PID control
plus limit control

Control output 1

15 amp @ 115V,
15 amp @ 230V

Control output 2

0.25 amp

Operating voltages

115 and 230 VAC

Maximum switching
load

15 amp @ 115 VAC
15 amp @ 230 VAC