Martin Christ Epsilon 2-10D LSCplus User Manual
Page 74

Freeze-dryer EPSILON 2-10D LSCplus
6 Operation
74
Version 04/2013, Rev. 1.5 of 16/12/2014 • sb
Translation of the original operating manual
Pos: 157 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50- 003 2-0 011 Prog ra mm erstell en Epsilo n @ 25\ mod _14 0498 399 771 8_6 8.do cx @ 1 839 70 @ 4 @ 1
6.5.3.1 Creating a program
To create a new program:
• Press the button "New program" in the main window "Program". The
dialog box that is displayed offers various different program templates.
The scroll bar on the right-hand side can be used to scroll through the
list.
1
Button for creating a
blank program
2
Scroll bar for
selecting a program
template
Fig. 65: Dialog box "Select program template"
Creating a blank program (1)
This button is used to call up a blank program template. Only section 1 is
pre-defined as the loading section. In this phase, the start conditions of the
programs are defined. The room temperature (20°C) is the standard
default.
NOTE
If the product is frozen outside the freeze-dryer (double-chamber method),
the shelf temperature must be adapted accordingly in section 1.
For all the other sections, the set values must be defined (see figure
below):
• Press the button “Insert section” (4) and select the position of the
section as well as the freeze-drying phase. The section will be inserted
at the defined position.
• Adapt the parameters of the inserted section. All of the values are not
available in all of the freeze-drying phases.
• Insert and edit the next section.
NOTE
During the creation of a program, the order of the individual freeze-drying
phases must be maintained. This means that it is not possible to insert a
freezing section after a main drying section.