64 Maintenance and Service

Operating Manual Freeze Dryer BETA 1-8 LDplus / BETA 2-8 LDplus

7.1.9. Checks by Operator

The operator has to ensure that the important parts of the freeze-
dryer relevant for safety are not damaged.
This especially refers to:

Lid or drying chamber
Seals
Oil level of vacuum pump
Accessories, especially changes like corrosion, wear

and tear of material etc.

Furthermore, an earth conductor check must be carried out
regularly.

7.2. Service

DANGER ! In the event of service work that requires
the removal of the panels, there is a risk of electric
shock or mechanical injury. Only qualified specialist
personnel is authorised to perform this service work.

The freeze-dryer is subject to high mechanical stress. In order to
be able to withstand this high level of stress, high-quality
components were used during the production of the freeze-dryer.
Nevertheless, wear cannot be excluded and it may not be visible
from the outside.
This is why we recommend having the freeze-dryer checked by
the manufacturer during an inspection once per year.

Information and appointments:
in Germany:
use the service request form at

www.martinchrist.de

[Service Area]

or contact

Martin Christ Gefriertrocknungsanlagen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Phone +49 (0) 55 22 / 50 07-84 25
Fax +49 (0) 55 22 / 50 07-94 25
E-mail: service@ martinchrist.de

outside Germany:
contact our agency in your country. All agencies are listed at

www.martinchrist.de

[Contacts] [Foreign agencies]

NOTE!

If you would like to utilise our service, please state the type of

your freeze-dryer and its serial number

Make use of our service request on the Internet. Please use

the request form (see above) on our website.

This manual is related to the following products: