5.7.1.1.
Set values for freezing
Timer
Time for the freezing phase section time. You can set a 
time between 1 minute and 200 hours. Selecting the 
infinite set value ( ) deactivates the timer for the 
freezing phase. 
5.7.1.2.
Set values for main drying & final drying
Timer
Time for the main or final drying phase section time. 
You can set a time between 1 minute and 200 hours. 
Selecting the infinite set value ( ) deactivates the timer 
for the main or final drying phase. 
Vacuum
Set value for the control vacuum in the drying chamber 
during main or final drying. The vacuum control option 
must be installed. You can set a value between 6.1 
mbar and 0.0010 mbar converted into steps of 1°C 
according to the vapor pressure curve above ice for 
water. 
Set values for final drying are not available unless the vacuum 
control option is installed. 
5.7.2. Special functions
The special functions menu is used to operate special accessories 
(defrosting device and electrical lifting device) of the freeze-dryer. 
If one of the options is not installed in the freeze-dryer, the system 
displays the message Option not available. 
A detailed description of the special functions can be found in the 
chapter “LDplus Special Functions”. 
 
5.7.3. Process and equipment information system
Further information can be found in the chapter “Process and 
equipment information system”. 