Rubber valves – Martin Christ Beta 2-8 LDplus User Manual
Page 41

Operating Manual Freeze Dryer BETA 1-8 LDplus / BETA 2-8 LDplus
Installation and Commissioning of the Unit 41
4.8.4. Rubber valves
The rubber valves (part no. 121860) enable the connection of 
round-bottom flasks, wide-neck filter bottles, or distributors for 
ampoules to a manifold or drying chamber. Depending on the 
connector of the components, the blue plug can be removed. 
1 Locking bolt
2 Connection to freeze-
dryer (e.g. via a manifold)
3 Aeration connection
4 Vessel connection
5 Rubber plug
 
 
 
NOTE! The rubber valves come supplied in an ungreased state. 
This is why a thin layer of vacuum grease must be applied to the 
connector of the freeze-dryer as well as to the vessel connector 
prior to start-up in order to ensure trouble-free operation. 
In position (1), the aeration connector is open and the vessel 
connector is closed. The accessory will be aerated while the 
vacuum inside the drying chamber is maintained. As a result, 
vessels can be exchanged without any interruption of the drying 
process. 
1
2
3
4
5
(1)
(2)
(3)