Storage – Labconco Biological Safety Cabinets 34269 Series User Manual
Page 57

Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
51
Original instructions
Resetting a Circuit Breaker:
To reset any of the circuit breakers located on the left side of the electronics
module, depress the white button until it sets.
Circuit Breakers- The front
breaker protects the electrical
outlets, the rear breaker protects
the motor and lights.
Storage
If the biosafety cabinet is to be left unused for more than one month, it should
be prepared for storage.
Note: The cabinet should not be stored in areas of excess humidity or
temperature extremes. If the cabinet is moved during storage, it must be
recertified before use.
1. Close the sash completely and seal the bottom edge and the exhaust outlet
with plastic sheeting.
2. Unplug the cabinet.
3. Ensure that the cabinet will not be moved or disturbed while being stored.
Figure 6-5
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series