Labconco Biological Safety Cabinets 34269 Series User Manual
Page 54
Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
48
Original instructions
Towel Catch Removal:
Although not normally required, the towel catch can be removed for cleaning,
inspection, etc.
Note: The work surface of the cabinet and the towel catch must be
thoroughly decontaminated before removing either.
1. Remove the work surface as described above.
2. Remove the towel catch by pivoting the bottom out toward you, as shown
in Figure 6-2; it is spring loaded, and you will feel some resistance. Note
the orientation of the towel catch.
3. Surface decontaminate the towel catch before removing it.
4. Reinstall the towel catch by sliding it back into position, in the correct
orientation. Also ensure that the Sampling/Decontamination Tube (the
black tube with an orange cap) rests behind the towel catch, allowing the
catch to contact the rear wall of the cabinet.
Sampling/Decontamination
Tube
Towel Catch Retaining
Spring
Towel Catch
Note: Upper and lower
towel catch flanges point to
the front of the cabinet, and
the bottom of the towel
catch rests against the back
wall.
Figure 6-2
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series