Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
49
Original instructions
Font Panel Removal and Installation:
1. Locate and remove the two Phillips
screws that secure the front panel as
shown in Figure 6-3. They are
located on the bottom corners of the
front dress panel.
2. Swing the bottom of the dress panel
out to clear the fluorescent light and
then lift the front dress panel
straight up and away from the
cabinet.
1. To reinstall the panel, reverse
these steps, ensuring that the
plastic pin in the top corners of
the dress panel properly engage
the corner posts.
Figure 6-4
Figure 6-3
This manual is related to the following products:
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series