Navigating the logic+ menu screens – Labconco Biological Safety Cabinets 34269 Series User Manual
Page 40
![background image](/manuals/738617/40/background.png)
Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
34
Original instructions
Navigating the Logic+ Menu Screens
MyLogic
TM
allows you to use the Smart-Start or Night-Smart features that
activate functions automatically when the sash is opened or closed.
If equipped, the UV lamp can be programmed to operate for a given time
interval when the sash is closed, before it shuts off.
NOTE: When you are in the menu mode, if a selection is not made within
30 seconds, the display will reset back to display mode.
Keypad operations are shown as
blue bold italic
. Menu screen selections are
shown as
green italics
.
NOTE: Pressing the appropriate touchpad button will override Smart-
Start or Night-Smart selections.
To access the menu, press the
Menu
button. The display panel will show the
first level menu. To select from the various menu options press the
▲
or
▼
buttons until the selected option is displayed. Press
OK
to accept that option,
or press
Menu
to return to the previous menu level.
Menu
Display
Mode Menu
Mode
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series