Directional Airflow
Directional airflow also plays a key role in biosafety cabinet performance.
Air is drawn into the front of the cabinet at the front grille. This “curtain” of
air makes it more difficult for aerosols to escape out of the work area of the
cabinet and into the outside environment. This airflow is often calculated and
referred to as the Inflow Volume or Average Inflow Velocity. This is
illustrated in Figure 4-3.
Figure 4-3
This manual is related to the following products:
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series