The common service operations necessary to maintain the biosafety cabinet
for peak performance are listed below.
Note: This manual covers operation and maintenance operations for the
owners/users of the Logic+ biosafety cabinets. Complete certification
procedures, service operations and specifications are published in a separate
publication Technical Manual: Purifier
®
Logic
®
+ Biosafety Cabinets. This
manual is available from Labconco’s website:
www.labconco.com
. A
complete certifier service kit is available to qualified certifiers from
Labconco. Call Labconco’s Product Service Department at 800-821-5525 or
816-333-8811.
Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.
Routine Maintenance Schedule
Weekly
• Wipe down the interior surfaces of the cabinet with a suitable
disinfectant, or 70% ethanol, and allow to dry.
• Using an appropriate glass cleaner, or LabSolutions
TM
Glass &
Surface Wipes, Labconco part # 1570000; clean the sash and the
surface of the UV lamp, if so equipped.
• Operate the cabinet blower, noting the percent filter life remaining in
an operational log.
Monthly (or more often as required)
• Using a damp cloth, or LabSolutions Glass & Surface Wipes,
Labconco part # 1570000; clean the exterior surfaces of the cabinet,
- Biological Safety Cabinets 33269 Series Biological Safety Cabinets 32269 Series Biological Safety Cabinets 34259 Series Biological Safety Cabinets 33259 Series Biological Safety Cabinets 32259 Series Biological Safety Cabinets 34249 Series Biological Safety Cabinets 33249 Series Biological Safety Cabinets 32249 Series Biological Safety Cabinets 34239 Series Biological Safety Cabinets 33239 Series Biological Safety Cabinets 32239 Series