Motor/Blower

The motor/blower assembly pulls room air into the front of the cabinet, and
re-circulates it internally. During its recirculation, the air is split into two
separate streams. One path leads through the exhaust HEPA filter and out of
the unit. The second path flows through the supply HEPA filter, which then
flows down through the work area, as shown in Figure 4-4. The motor in the
Logic+ cabinet is an electronically commutated motor (ECM). The ECM is a
brushless DC motor that includes its own power supply to convert the
incoming alternating current to direct current, as well as its own
microprocessor to control and measure the motor’s operation. The motor
utilizes Labconco’s exclusive Constant Airflow Profile

TM

(CAP) program to

deliver a consistent volume of air, throughout the life of the cabinet.





















Figure 4-4