beautypg.com

Hanna Instruments HI 83200C User Manual

Page 61

background image

120

121

10 mL

• Add the content of one packet of HI 93751-0

Indicator reagent.

• Replace the cap and invert gently to mix for

1 minute (about 30 inversions).

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in

mg/L of Sulfate (SO

4

2

¯).

INTERFERENCES

Interferences may be caused by:
Calcium (as CaCO

3

) above 20000 mg/L

Chloride (as Cl

-

) above 40000 mg/L

Magnesium (as MgCO

3

) above 10000 mg/L

Silica (as SiO

2

) above 500 mg/L

Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous
filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.

Sulfate

12

3

6

9

1’

SPECIFICATIONS

Range

0 to 100 mg/L

Resolution

5 mg/L

Accuracy

±5 mg/L ±3% of reading at 25 °C

Light Source

Tungsten lamp with narrow band interference filter @ 466 nm

Method

Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension
is measured.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93751-0

Indicator reagent

1 packet

REAGENT SETS

HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure on page 17.

• Select the

Sulfate method using the procedure described

in the

Method Selection section (see page 12).

• Fill a cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

Sulfate

SULFATE