Hanna Instruments HI 83200C User Manual

Page 24

background image

46

47

Color of Water

• Fill the second cuvette up to the mark with unfiltered sample

and replace the cap. This is the apparent color.

• Filter 10 mL of sample through a filter with a 0.45 µm

membrane into the third cuvette, up to the 10 mL mark and
replace the cap. This is the true color.

• Insert the apparent color cuvette (# 2) into the instrument

and close the lid.

• Press

Read to start the reading.

• The meter displays the value

of apparent color in PCU.

• Remove the cuvette, insert the true color cuvette (# 3) into the

instrument and ensure that the notch on the cap is positioned
securely into the groove.

• Press

Read to start the reading. The meter displays the value of true color in PCU.

Color of Water

COLOR OF WATER

SPECIFICATIONS

Range

0 to 500 PCU (Platinum Cobalt Units)

Resolution

1 PCU

Accuracy

±10 PCU ±5% of reading at 25 °C

Typical EMC

± 1 PCU

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Colorimetric Platinum Cobalt method.

REQUIRED ACCESSORIES

0.45 µm membrane for true color measurement.
For other accessories see page 128.

MEASUREMENT PROCEDURE

• Select the

Color of Water method using the procedure described in

the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with deionized water and replace

the cap. This is the blank.

• Place the blank (# 1) into the holder and close the lid.

• Press the

Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank.

# 1

# 1

10 mL

# 3

# 2

10 mL

# 2

# 3