F R E E C H L O R I N E
F R E E C H L O R I N E
F R E E C H L O R I N E
F R E E C H L O R I N E
F R E E C H L O R I N E
SPECIFICATIONS
Range
0.00 to 2.50 mg/L
Resolution
0.01 mg/L
Accuracy
±0.03 mg/L ±3% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
EPA DPD method 330.5. The reaction between free chlorine and the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER
:
Code
Description
Quantity
HI 93701-0
DPD
1 packet
LIQUID:
Code
Description
Quantity
HI 93701
A-F
DPD1 Indicator
3 drops
HI 93701
B-F
DPD1 Buffer
3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Free Chlorine method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Free Chlorine
FREE CHLORINE
• Press
Timer or wait for 5 minutes. Then invert again
the cuvette 10 times to mix (about 15 seconds).
• Reinsert the cuvette into the instrument.
• Press
Read to start the reading. The instrument displays the results in mg/L of Calcium.
INTERFERENCES:
Interferences may be caused by:
Acidity (as CaCO
3
) above 1000 mg/L
Alkalinity (as CaCO
3
) above 1000 mg/L
Magnesium (Mg
2+
) above 400 mg/L
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when
the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Using the 1 mL syringe, add exactly 1 mL of the
HI 93752B-0 Calcium Oxalate Reagent. Replace the
cap and invert the cuvette 10 times to mix (about
15 seconds).
Calcium
10 mL