Hanna Instruments HI 83200C User Manual
Page 40

78
79
Manganese LR
• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent
solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.
• Add 1 mL of the HI 93748C-0 0.1% PAN indicator
solution to each cuvette, replace the caps and shake
gently.
• Add 3 drops of Dispersing Agent (HI 93703-51) to each
cuvette, replace the caps and invert gently to mix for
about 30 seconds.
• Place the cuvette with the reacted deionized water (blank)
into the holder and close the lid.
• Press
Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for
2 minutes and then press
Zero. The display will show "-0.0-" when the meter is zeroed and ready for
measurement.
Manganese LR
MANGANESE LOW RANGE
SPECIFICATIONS
Range
0 to 300
µ
g/L
Resolution
1
µ
g/L
Accuracy
±10
µ
g/L ±3% of reading at 25 °C
Typical EMC
±1
µ
g/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm
Method
Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93748
A-0
Ascorbic acid
2 packets
HI 93748
B-0
Alkaline-cyanide sol.
0.40 mL
HI 93748
C-0
0.1% PAN indicator
2 mL
HI 93703-51
Dispersing Agent
4-6 drops
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Manganese LR method using the procedure described in the Method Selection section (see
page 12).
• Fill one cuvette with 10 mL of deionized water (up to the mark).
• Fill a second cuvette with 10 mL of sample (up to the mark).
• Add the content of one packet of HI 93748A-0 Ascorbic acid to
each cuvette, replace the caps and shake gently until completely
dissolved.
10 mL
10 mL