Hanna Instruments HI 83200C User Manual
Page 35

68
69
Iron HR
IRON HIGH RANGE
SPECIFICATIONS
Range
0.00 to 5.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.04 mg/L ±2% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
EPA Phenantroline method 315B, for natural and treated waters.
The reaction between iron and reagents causes an orange tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93721-0
Powder Reagent
1 packet
REAGENT SETS
HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Iron HR method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI 93721-0 reagent. Replace the cap and shake until
dissolution is complete.
Iodine
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press
Read. When the timer ends the meter will perform the
reading.
• The instrument displays concentration in
mg/L of iodine.
INTERFERENCES
Interference may be caused by: Bromine, Chlorine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately 2
minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
10 mL