Hanna Instruments HI 83200C User Manual
Page 41

80
81
Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range
0.0 to 40.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.3 mg/L ±5% of reading at 25 °C
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93730
A-0
Reagent A
1 packet
HI 93730
B-0
Reagent B
1 packet
HI 93730
C-0
Reagent C
1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Molybdenum method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Manganese LR
• Insert the second cuvette with the reacted sample into the instrument.
• Press
Read to start the reading. The instrument displays the results in
µµµµµ
g/L of manganese.
• Press ▲ or ▼ to access the second level functions.
• Press the
Chem Frm key to convert the result in
µ
g/L of potassium permanganate (KMnO
4
) and
permanganate (MnO
4
).
• Press ▲ or ▼ to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
3
Nickel above 40 mg/L
Zinc above 15 mg/L.
10 mL